发布时间:2019-04-25 09:12 原文链接: DynamicFlowAssayinaParallelPlateFlowChamber

Dynamic Flow Assay in a Parallel Plate Flow Chamber


John T. Patton~GlycoTech Corporation, Rockville, Maryland 20850


Flow assays allow visualization of cell adhesion under well-defined wall shear stress. The visualization of the different events of cell adhesion can be quantified by selective image acquisition and subsequent image processing. Flow assays are uniquely suited to the investigation of adhesive events which occur very rapidly in a time scale shorter than that of most static adhesion assays. In addition, events subsequent to the initial events can be studied such as cell stabilization and spreading giving some insight into the kinetics of particular cell-cell or cell-substrate adhesive behavior.

 

Materials:

Microscopy:

Inverted-stage microscope configured for phase contrast and fluorescence operation

Objective lens (6.3 x, 10 x, 40 x)

Stage incubator

Biological:

Cell suspension (neutrophils, cancer cells)

Cell monolayer or coated substrate in 35 mm dishes

Adhesion media (culture media serum-free with 12 mM Hepes)

Fibronectin (human plasma)

Bovine serum albumin (BSA)

Flow Chamber System:

Flow chamber deck with gasket and tube fittings

35 mm tissue culture dishes

Harvard syringe pump

Glass syringes to fit on pump (3, 5, 20, or 60 cc)

Silastic tubing

Stopcocks

Vacuum pump

Computer/Electronics:

Digital image acquisition and processing system

Display monitor

Video recorder

CCD camera with controller

Time-date generator

Computer with interface to image system

Storage device for images

 

 

Protocol:

Cell Monolayer Preparation

1. Coat dishes with human fibronectin (FN) by adding 1 ml of 5.5µg/ml FN solution to each dish.

2. Incubate for 30 min. at room temperature.

3. Aspirate solution out of each dish.

4. Add 2 ml of cell suspension (Huvecs, CHO cells) to each dish at seeding density of 0.5-3.0 x 105 cells/ml depending on time required to reach confluence.

5. After 1-5 days, use dishes with confluent monolayer in flow assay. Feed cells every 2-3 days, but usually not with 48 hrs. of flow assay.

Coated Substrate Preparation

1. Outline a coating region (5 mm diameter) for coating substrate in the center of each dish with marking pen.

2. Add 20µl of solution containing substrate at a concentration of 10µg/ml to coating region. Incubate 1 hr.

3. Aspirate off liquid, add 20µl of 1% BSA to block for 1 hr.

4. Aspirate off BSA, add 20µl of inhibitor for 1 hr.

5. Dishes are ready for use in flow assay.

Flow Assay Using Parallel Plate Flow Chamber

1. Turn on stage incubator and warm adhesion media to 37°C 1 hr. prior to beginning flow assay.

2. Assemble flow system apparatus connecting inlet, outlet, and vacuum lines to the flow chamber deck. Fill system with media and remove all air from system.

3. Fill inlet reservoir with cell suspension. For assays using cell monolayers, 106 cells/ml is recommended. For coated substrate assays, use 105 cells/ml.

4. Attach dish to flow chamber deck by holding the deck inverted, place a small bubble of media on flow path area, then place 35 mm dish on the deck. Vacuum will hold dish on deck. Make sure dish was attached with no air bubbles in the flow path.

5. Place assembled chamber on microscope stage.

6. Initiate flow of cells syringe pump connected to outlet flow chamber at a shear stress in the range of 0.1-4.0 dynes/cm2.

7. Allow cells to flow for sufficient time to get an adequate number of cells interacting with the cell monolayer or coated surface. Generally 3-10 min. is used.

8. Begin image acquisition. Collect images at 7-10 locations on the dish. Generally 3 dishes at a given experimental condition gives enough data to show statistical differences between treatments.

9. After images are acquired on all dishes, perform image analysis to quantify the flow assay.

References:

(1) Lawrence, M.B., McIntire, L.V., Eskin, S.K., (1987), Effect of flow on polymorphonuclear leukocyte/ endothelial cell adhesion, Blood 70: 1284-1290.

(2) Patton, J.T., Menter, D.G., Benson, D.M., Nicolson, G.L., McIntire, L.V., (1993), Computerized analysis of tumor cells flowing in a parallel plate chamber to determine their adhesion stabilization lag time, Cell Motility and the Cytoskeleton 26: 88-98.

(3) Jones, D.A., Abbassi, O., McIntire, L.V., McEver, R.P., Smith, C.W., (1993), P-selectin mediates neutrophil rolling on histamine-stimulated endothelial cells, Biophysical Journal 65: 1560-1569.


相关文章

安捷伦宣布推出AssayMAPBravo蛋白质样品前处理Workbench4.0

将 AssayMAP 的使用范围扩展至法规监管环境2022年11月28日,北京——安捷伦科技有限公司(纽约证交所:A)近日宣布推出AssayMAPBravo蛋白质样品前处理Work......

“3CoinONE全新体验气袋进样器”ANTOP奖进入专家评审阶段

万物蓬勃的7月里迎来了2022年ANTOP奖的申报和评审工作。由岛津企业管理(中国)有限公司申报的“3CoinONE全新体验气袋进样器”ANTOP奖进入专家评审阶段。奖项名称:3CoinONE全新体验......

青源峰达太赫兹在国际顶级期刊《TrendsinBiotechnology》发文

青岛青源峰达太赫兹科技有限公司研发团队在国际顶级期刊《TrendsinBiotechnology》(译名:《生物技术趋势》)在线发表题为“THzmedicalimaging:frominvitroto......

firstinclass胃癌一线疗法获FDA突破性疗法

安进宣布,美国FDA授予其在研firstinclass单抗bemarituzumab突破性疗法认定,与改良FOLFOX6化疗方案(亚叶酸钙、氟尿嘧啶和奥沙利铂)联用,一线治疗FGFR2b过表达和HER......

金属所三维伊辛模型精确解研究取得进展

2007年,中国科学院金属研究所研究员张志东在英国《哲学杂志》(PhilosophicalMagazine)上发表论文,提出两个猜想,并在猜想基础上推定出三维伊辛模型的精确解。被《哲学杂志》审稿人评价......

硬核原创FlowLab:化工制药业绿色安全生产的法宝

分析测试百科网讯,今年来化工厂的频繁爆炸事件造成极大社会反响,化工生产中的安全问题再次触动了所有国人的神经。安全排查、人员责任等手段固然能起到一定的效果,但还不能从根本上解决问题;分析诸多事件,我们发......

采用USEPA方法OIA1677和FlowSolution®IV测量有效的氰化物

这里描述了按照USEPA方法OIA-16771-4,通过价位交换、流动注射分析(FIA)和安培计检测有效氰化物的方法。这个方法用于USEPA关于清洁水法案,资源保护与回收法案,全面环境响应,补偿与责任......

新一代蛋白结构预测技术

《NatureChemicalBiology》杂志在线版一篇文章介绍,伊利诺斯州立大学JohnA.Gerlt博士率领的研究小组研制出一种新途径,能够确定氨基酸序列已知的蛋白的结构和功能。这是首次利用计......

微孔板封膜穿孔机PinPlate选择指南

......

精品:GeneFlowfromGMPlants+好书一本

今天在网上淘到的一本书GeneFlowfromGMPlants,05年,还算比较新。望大家多多支持。GeneFlowfromGMPlantsEditedbyGUYM.POPPYSchoolofBiol......