1) Turn on the counter by pulling out the on/off button. You need to do this at least 10 min before use to obtain sufficient vacuum.
Usually put 0.2 ml of your suspension of cells into 10 ml Tris or TD. In reality you can use any dilution, as long as you know how to multiply.
2) Be sure to mix the sample thoroughly before counting, since the cells will settle to the bottom. To count the sample, turn the glass valve directly above the cells to vertical (open). Wait until you see a green baseline in the "sizing" monitor--the vertical "blips" represent either cells or crud. The larger the blip, the larger the cell. Turn valve A back to horizontal (closed) to begin the count.
The length of the count should be approx. 13 sec, from when the first number appears in the register to when the green display goes out. If the counting time is greater than 13 sec, the aperture is clogged with debris and should be cleaned gently with the brush provided (as well as "flushed," see 5). Any count which is obtained when the aperture is obstructed or downright clogged has an unpredicatable coincidence and should be done again.
The count displayed is the the number of cells in 0.5 ml. If you diluted 0.2 ml of cells to 10 ml, multiply the number displayed by 100 to determine the number of cells per ml in the original culture. Be sure to use the coincidence correction table if your count is greater than 10,000. At high concentrations the probability of two or more cells passing through the aperture simultaneously, and being counted as one cell, is high.
Find the first three digits of your count on the table and use the corrected figure, which is next to it, for your calculations. Obviously, if the cell count displayed is too low or too high (less than 3000 or greater than 99,900 cells per 0.5 ml), you should set up an appropriate dilution and re-count the sample.
3) When you're finished counting, flush the system. Put the electrode into the vial containing Isoterge and check to see that the flask labeled "Rinse Flask" contains at least 100 ml of solution. If it's low, fill it with Tris or TD and some Isoterge. Turn valve A to vertical, then turn valve B, which is off to the left, to vertical. Flushing will occur. After 10 sec, turn valve B back to horizontal and then turn valve A to horizontal. Store the electrode in the vial of Isoterge, and, if it's the end of the day, turn the machine off.
近日,服务科学领域的全球领导者赛默飞世尔科技(以下简称赛默飞)宣布,在达成收购意向两个月之后,赛默飞以28亿美元、折合人民币约190亿元的价格,完成了对TheBindingSiteGroup的全现金收......
11月15日,施普林格·自然和TheLens平台宣布结成重要的合作伙伴关系,以更深入地揭示学术研究和数据如何能通过经济和社会成效,加速推动创新的问题解决方式。通过将科学、投资和企业领域的开放数据更好地......
一次性反应技术近年来逐渐成为生物制药企业、CDMO公司的选择,由于其污染风险少、清洁、验证成本低的特点,备受业内好评。一次性技术选择不仅仅为生物制药企业和CDMO公司带来一种创新方向的选择,更重要的是......
6月10日,QS教育集团正式发布了2021年世界大学排名,中国共有83所高校上榜,包括内地高校51所,港澳台地区高校32所。中国大学的总体排名情况已经连续数年呈上升趋势,今年再度刷新了榜单。大学排名,......
自然杀伤细胞(Naturalkillercells,NKcells)是一种效应性淋巴细胞,在人体抗肿瘤过程中发挥重要作用,但是,仍然有一些肿瘤细胞能够逃避机体NK细胞的攻击,逐步发生发展,侵袭转移,严......
作为一种重要的植物激素,茉莉酸不仅调控植物对于机械损伤、昆虫取食和腐生型病原菌侵害的防御反应,还参与调控诸多生长发育过程。basicHelix-Loop-Helix(bHLH)类型转录因子MYC2是茉......
ThePlantCell是植物领域的著名学术期刊,对植物学的发展起到了重要的引领作用。为庆祝创刊30周年,ThePlantCell杂志社邀请部分编委会成员及其他科学家对发表在该杂志的重要研究工作进行评......
研究人员发现了小鼠体内一种新的转录因子可以帮助调节间充质干细胞(mesenchymalstemcells,MSCs)分化为骨的过程。目前科学家们对于骨细胞分化的研究并不深入,而MSCs则是再生医学领域......
分析测试百科网讯近日,全球领先的基因组学服务提供商——GENEWIZ已经签署收购BeckmanCoulter基因组学服务业务的最终协议,该业务提供医疗保健、学术和政府机构的专业的基因组测序解决方案(G......
分析测试百科网讯2015年7月16日,丹纳赫间接全资附属公司——BeckmanCoulter宣布,该公司已经达成一项最终协议,收购西门子医疗诊断的临床微生物学业务。西门子临床微生物学业务是在微生物鉴定......