Each time the microscope is to be used it should be set up correctly to give a good image. Most often users forget to adjust the iris diaphragm and focus the condenser to give its optimum performance, this could result in objects in the preparation being undetected.
There are many microscope set up procedures for specific functionality but the most versatile for general use is ‘Critical Illumination’ as detailed below.
Examine the microscope and identify the parts as siting of some controls may vary and some have ‘Fixed’ or specialised condensers.
Ensure the light intensity control is set to low, turn on the light and increase the brightness to give a white light output.
Raise the substage condenser fully and open the iris diaphragm fully.
Rotate the nosepiece to the lowest power objective (x4).
Rack the stage down using the coarse adjustment control and place the specimen slide on the stage using the retaining mechanism to hold in place.
While looking carefully rack the stage up close to the objective lens (do not rack the lens into the slide; this cannot normally occur when a low power objective is selected).
Set the eye pieces to the correct width for your eyes so you can look down both together comfortably.
Focus the Microscope.
Look down the eyepieces with both eyes open and slowly lower the stage to bring the specimen into focus, (use the coarse adjustment control initially and then the fine). Slight movement of the specimen while focusing by using the mechanical stage controls may aid you to find the object. If the image is not sharp focus adjustment of the eyepieces may be necessary. If one eyepiece is fixed and one adjustable, close the eye over the adjustable one. Now focus using the fine control until a sharp image is obtained, (if you cannot achieve this ask for assistance as the problem may lie elsewhere). Now open the other eye (keep both open) and using the adjustment on the eyepiece itself rotate to obtain a sharp image.
Focus the Condenser.
The condenser should be fully raised to start with. Rotate the nose piece to select the x10 objective and check the focus using fine adjustment. Position the point of a pencil into the middle of the light source over the lamp housing. Using the condenser control, slowly rack it down until the outline image of the pencil becomes sharp and crisp, (this is the ‘Critical’ point for the condenser for optimum performance from where the technique derives it’s name). Note this must be done only after a specimen has been focused on the stage.
Set the Iris Diaphragm.
When the iris diaphragm is fully open the image is flooded with light and definition is lost due to ‘white-out’. As the diaphragm is closed controlling the amount of light passing through the condenser the image is darkened and outlines appear thickened and defined as more contrast is achieved. Remove one eyepiece and place safely to one side. Close the diaphragm down ,then open slowly until you achieve a dark ring blanking off about 1/5th of the original field of view to give a ‘bicycle tyre’ effect. The correct amount of light should thus be obtained.
Examine the specimen, moving it by means of the mechanical stage controls. No further adjustment to the condenser or iris diaphragm is usually necessary to view other slides of the same type/density at low power. The Vernier scale readings may be used to identify and re-locate a point of interest on a slide provided it is a ‘mounted’ preparation and the slide is placed on the stage in the same orientation, i.e. well labeled.
日前,国产期刊TheInnovation获得首个影响因子(IF=32.1),成为科睿唯安JCR综合性期刊分类下排名仅次于《自然》(IF=64.8)和《科学》(IF=56.9)的期刊,并且这本期刊在目前......
现在,美国的消费者可以预订一种持续发光的基因工程植物了。美国48个州的居民只需花29美元就可以得到一株能持续发出淡淡绿色光芒的矮牵牛。美国生物技术公司LightBio将于4月份开始运送一批5万株“萤火......
近日,服务科学领域的全球领导者赛默飞世尔科技(以下简称赛默飞)宣布,在达成收购意向两个月之后,赛默飞以28亿美元、折合人民币约190亿元的价格,完成了对TheBindingSiteGroup的全现金收......
11月15日,施普林格·自然和TheLens平台宣布结成重要的合作伙伴关系,以更深入地揭示学术研究和数据如何能通过经济和社会成效,加速推动创新的问题解决方式。通过将科学、投资和企业领域的开放数据更好地......
一次性反应技术近年来逐渐成为生物制药企业、CDMO公司的选择,由于其污染风险少、清洁、验证成本低的特点,备受业内好评。一次性技术选择不仅仅为生物制药企业和CDMO公司带来一种创新方向的选择,更重要的是......
蛋白质作为构成人体组织器官的支架和主要物质,在人体生命活动中起着重要作用。蛋白质的相互作用能产生许多效应,如形成特异底物作用通道、生成新的结合位点、失活、作用底物专一性和动力学变化等,细胞的代谢、信号......
2021年9月9日,无锡臻和生物科技有限公司(以下简称“臻和科技”)与美国VyantBio公司签署TissueofOrigin®(以下简称“TOO®”)全球权益和ZL转让协议,全资收购这款唯一获FDA......
2021年7月20日,JournalofCellularPhysiology及JournalofCellularBiochemistry同时撤回了中国学者49篇文章。从2019年开始,Journalo......
6月10日,QS教育集团正式发布了2021年世界大学排名,中国共有83所高校上榜,包括内地高校51所,港澳台地区高校32所。中国大学的总体排名情况已经连续数年呈上升趋势,今年再度刷新了榜单。大学排名,......
磷酸甘油酸突变酶1(PGAM1)通过其代谢活性以及与其他蛋白质(例如α平滑肌肌动蛋白(ACTA2))的相互作用,在癌症代谢和肿瘤进展中起关键作用。变构调节被认为是发现针对PGAM1的高选择性和有效抑制......