A. Gilson-style Pipetmen Micropipets (and similar models)


We primarily use the Gilson micropipets in the core course labs. We have four sizes identified by the number on the round button on the plunger. The value is the maximum volume in microliters that can be transferred with that size pipet.


 

Size indicator button on a Gilson P1000.

 Other autopipet sizes we have on hand. P1000 P200 P100 P20


What size of micropipet is right for the job?

RULE OF THUMB: Always select the SMALLEST size pipet that will handle the volume you wish to move to achieve the greatest accuracy. Accuracy decreases as you use unnecessarily large pipets for small volumes.

The following table shows the useful volume ranges for each pipet type. Make special note of the min and max values.


 


 SIZE

 


 Useful Range

 


 Max Volume

 


 Min Volume


Gilson-style P1000;
Finnpipette 100-1000 ul


 ~200-1000 ul


 1000 ul


 200 ul


 Gilson-style P200;
Finnpipette II 20 -200 ul


 20-200 ul


 200 ul


 20 ul


 Gilson-style P100


10-100 ul


 100 ul


 10 ul


 Gilson-style P20;
Finnpipette II 2 -20 ul


 ~0.5-20 ul;
2 - 20 ul


 20 ul


 0.5 ul; 2 ul


 

B. How to Read the Volume on the Micropipet

Look at the front face of the pipet and you will see a window with 3 (three) digits inside. The diagram below shows the MAXIMUM value that can or should be dialed in on each size pipet. To exceed these values will put the pipet out of calibration. Beside each "window" below is the numbers place it represents. Please take the time to learn how to read them so as to avoid damaging them by dialing values out of their range.

Note that some of the smaller volume models have one or two decimal places.



 


Gilson-style P1000 scale window.
   P1000   P200,P100   P20   11000's  100's 10's 2,1100's 10's 1's  2 10's 1's 1 decimal  0  0  0  0  0  0


 

 

C. How to change the load volume:

Hold the pipetor horizontally with the plunger to your left. Change or set the load volume by turning the knurled plastic knob at the top of the hand grip. Roll it towards you to decrease the values and away from you to increase. WATCH CAREFULLY as you change the values to make sure you dont take the numbers out of range. If in doubt, review the allowable volumes for each size pipetor, and then ask for help.

 

D. Technique

  • GRIP: There are two typical ways that people hold micropipets: the index finger grip, and the thumb grip (see figure below). Hold the pipetor like a knife in a horror flick with your index finger OR thumb on the plunger. While it's a matter of personal preference, some argue that your index finger gives you much finer control over the plunger action. As seen in the photos below, however, some micropipets are clearly designed for the thumb grip.



 


 


 Index finger grip


 Thumb grip


  • LOADING: Load a sterile tip (blue for 1000 ul pipettes; yellow for P200 and all smaller sizes including) and then reclose the tip box to maintain sterilitiy.

    • Push the plunger down slowly to the point of first resistance: this is the load volume.

    • While holding the plunger at the load volume set point, put the tip into the solution so that it is immersed just enough to cover the end (3-4 mm).

    • Slowly release the plunger to draw up the liquid making sure to keep the tip immersed. Visually inspect the load to make sure it is correct - there should be no air space in the distal end tip.

    • Performance may be improved by prewetting the tip once or twice (load and discharge a small amount) before actually loading for delivery.

  • DELIVERY: To deliver the volume, place the tip into the receiving vessel and press the plunger all the way to the bottom - this expells all the liquid and gives a little extra volume to get the last droplet out - AND THEN, WITHOUT RELEASING THE PLUNGER, withdraw the tip.

  • CHANGE TIP? Repeat as necessary using the same tip if NOT changing solutions/cultures. Use a fresh tip for every change of solution, or whenever it is prudent to maintain sterile conditions and avoid cross contamination of solutions or cultures. 

  • TIP DISCHARGE: While holding the tip over an appropriate waste receptacle, press the white tip discharge slider on the back of the grip.


相关文章

回顾和预测现代生物制药为何青睐一次性生物反应器?

一次性反应技术近年来逐渐成为生物制药企业、CDMO公司的选择,由于其污染风险少、清洁、验证成本低的特点,备受业内好评。一次性技术选择不仅仅为生物制药企业和CDMO公司带来一种创新方向的选择,更重要的是......

科学家将人工智能技术成功用于蛋白质复合物结构预测

蛋白质作为构成人体组织器官的支架和主要物质,在人体生命活动中起着重要作用。蛋白质的相互作用能产生许多效应,如形成特异底物作用通道、生成新的结合位点、失活、作用底物专一性和动力学变化等,细胞的代谢、信号......

发力癌症分子病理诊断,无锡臻和全资收购TissueofOrigin®

2021年9月9日,无锡臻和生物科技有限公司(以下简称“臻和科技”)与美国VyantBio公司签署TissueofOrigin®(以下简称“TOO®”)全球权益和ZL转让协议,全资收购这款唯一获FDA......

这3个杂志撤回中国学者249篇文章,包含上交、中山等名校

2021年7月20日,JournalofCellularPhysiology及JournalofCellularBiochemistry同时撤回了中国学者49篇文章。从2019年开始,Journalo......

肿瘤治疗的强心剂,中国学者开发肿瘤治疗新策略

磷酸甘油酸突变酶1(PGAM1)通过其代谢活性以及与其他蛋白质(例如α平滑肌肌动蛋白(ACTA2))的相互作用,在癌症代谢和肿瘤进展中起关键作用。变构调节被认为是发现针对PGAM1的高选择性和有效抑制......

llumina宣布推出新型基因分型芯片|支持AllofUs研究计划

2018年12月6日,来自圣迭戈的消息——Illumina公司(纳斯达克股票代码:ILMN)今天宣布推出新型高密度基因分型芯片Infinium™GlobalDiversityArray。这款芯片设计源......

SDSPAGE异常电泳现象及分析SDSPAGEHallofShame

SDS-PAGE异常电泳现象及分析SDS-PAGEHallofShame.pdf  很不错的东东~~推荐下~......

Preparationofdenaturing6%

Preparationofdenaturing6%polyacrylamidegelsformicrosatelliteanalysis(alsoforSSAP,high-resolutionIRAP......