Introduction

Biofilms are microbes attached to a surface. The  microbes form a film on the surface, giving rise to the name biofilm. This Application Note deals specifically  with bacterial biofilms that convert chemicals to  electrical current on electrodes. Because of this   function, we refer to them as electrochemically active biofilms (EABs). Interest in EABs ranges from their function as bioanodes and biocathodes in microbial fuel cells to new types of biosensors as well as novel biosynthesis of sustainable chemicals.

This application note shows results for Geobacter sulfurreducens, an oxygen-intolerant species of bacteria able to grow on  electrodes.1

If you are not familiar with quartz-crystal microbalances, see our Application Note “Basics of a quartz crystal microbalance”.

Experimental Notes

Culturing of G. sulfurreducens biofilms is not discussed in this Application Note. G. sulfurreducens is a non- pathogenic  BSL-1  biological agent.

Unlike typical electrochemical experiments, current  must be collected over several days to give time for the biofilm to grow. Choose an appropriate sampling rate, such as one measurement every five minutes. We show only one day of  growth.

Gamry’s  jacketed  EuroCell™  Electrochemical  Cell  Kit was used as the reactor vessel. An eSorptionProbe (Part No. 971-18) with 10 MHz Au-coated crystals  was  used.  A saturated Ag/AgCl electrode was used as  reference.

All parts of the reactor vessel were sterilized prior to biofilm growth.2  The assembled reactor vessel  was therefore sterile. Biological growth medium was then added. Oxygen was purged from the vessel using a gas mixture  of  20%  CO2   and  80%  N2.  The  electrode potential was then fixed to 0 VAg/AgCl using Gamry’s Framework™ software with the chronoamperometry script. A suspended cell culture of G. sulfurreducens was added and current was  recorded.

Results

Basics of biofilm growth

Figure 1. Photograph of biofilm grown on the 10 MHz Au-coated crystal for the  eSorptionProbe.

Biofilm can be grown on the Au-coated crystal (Figure 1). The pink hue on the Au surface is the biofilm. The biofilm is only found on Au because the electrode is the electron sink for the electrons generated from the  biofilm metabolizing acetate (a source of organic  carbon). In order to survive and grow, the biofilm must have an electron sink. As a result, surfaces such as the plastic coating that do not accept electrons show no biofilm growth.

The biofilm acts as a catalyst for the electrochemical conversion of acetate to carbon dioxide. The half-reaction above is an overall description of what happens in an individual cell in the biofilm. It does not account for carbon assimilation as biomass. The exact path through which electrons follow to reach the electrode surface is complex and not discussed in this Application Note.

For G. sulfurreducens biofilm, the acetate half-reaction is activated at electrode potentials more positive than approximay −0.4 VAg/AgCl. This  measured  value  is  true for most Au and C surfaces. Figure 2 shows the result of fixing the electrode potential to 0 VAg/AgCl in the presence of G. sulfurreducens cells and acetate. Cells that attach to the electrode form the initial biofilm.To survive, the initial biofilm metabolizes acetate and produces electrons.This produces an electron flow and is the cause of the increase in current in Figure 2.

Figure 2. Current output over time of a growing G. sulfurreducens biofilm. The electrode potential was fixed at 0 VAg/AgCl.

As the biofilm grows, current increases. In effect, the current output over time for EABs is a growth curve. Exponential growth can be easily checked by plotting the data in Figure 2 on a semilogarithmic plot (Figure 3). The initial region is the pre-exponential growth phase or lag phase. The linear region is exponential biofilm growth.

Figure 3. Semilogarithmic plot of current output over time. The linear region reflects exponential biofilm growth.

Standard practice is to confirm existence of the biofilm   by using scanning electron microscopy. Figure 4 shows a micrograph of the biofilm after fixation (standard procedure). The flat background is the bare Au surface. The biofilm was purposely removed to show the Au surface and biofilm in the same image.  The micrograph   is a visual indication that biofilm produced the current observed. If the biofilm is distinct like the pink hue in Figure 1, a photograph may be   sufficient.

Figure 4. Scanning electron micrograph of the biofilm on the Au-coated crystal after growth and a standard fixation procedure. After fixation, the biofilm is no longer alive. Single cells and multi-layered cell clusters are visible.

Biofilm voltammetry

During biofilm growth, the chronoamperometry script  can be stopped without damaging the biofilm. This is valuable for a simple voltammetric sweep. Usually cyclic voltammetry is the preferred script because of its simplicity. Figure 5 shows two cyclic voltammograms of    a growing biofilm (replicate) at 44 hours and 49 hours of growth. A catalytic wave is observed with several redox peaks superimposed between −0.33 VAg/AgCl  to  −0.23 VAg/AgCl. As the scan reached positive potentials, the effect of the redox peaks is minimized. In this potential region, limiting current is observed. The height of the catalytic wave increases with biofilm  growth.

 

Figure 5. Cyclic voltammograms of a G. sulfurreducens biofilm during growth. Scan rate was 30 mV/s. (Note that the chronoamperometry scan was stopped in order to run the cyclic voltammetry script.)

QCM frequency shift

Cell attachment and biofilm growth can be monitored in real-time using Gamry’s electrochemical quartz-crystal microbalance (eQCM). Using the Gamry Resonator™ software, both series frequency-shift (dFseries) and current can be recorded simultaneously. Figure 6 shows dFseries and Reduced (secondary y-axis) during   biofilm growth. This was recorded simultaneously with the current shown in Figure 2. The graphs are separated for clarity.

Figure 6. dFseries and Reduced over time decreasing in response to biofilm growth on the QCM. These data were taken simultaneously with the current data from Figure 2.

During the time it took the biofilm to reach a current output  of  40  µA  in  Figure  2,  dFseries  reached  −1250 Hz. Usually, for rigid films, dFseries can be converted to mass using the Sauerbrey Equation. However, the decrease in Reduced from 1500 to 1000 indicates a significant decrease in the rigidity of the biofilm. As a reference, a −1800 Hz dFseries shift during copper plating onto the QCM results in a minimal decrease in Reduced from 1280 to 1240. The important point is that it is typically incorrect to use the Sauerbrey  Equation to convert dFseries to mass for a biofilm, because of the biofilm’s viscous nature.

For most rigid film depositions using the eQCM, a plot of dFseries vs charge yields a straight line (for example, copper plating). In the case of G. sulfurreducens biofilm, this is not true.2 The reason is that charge passed does  not determine how much biofilm has grown on the electrode surface. For example, it is possible to have the total charge passed at the electrode increase with time without biofilm growth. Current passed over time is  more relevant for G. sulfurreducens biofilm because current is a proxy for biofilm growth. Figure 7 confirms this expectation and shows a linear correlation between dFseries and current.

Figure 7. dFseries vs. current (linear fit in   red).

Conclusions

This application note introduces electrochemically active biofilms to researchers outside of the field, and shows how the eQCM can be integrated into basic electrochemical techniques. It may also be useful to high school, undergraduate, or new graduate students who want to learn more about the techniques used to study electrochemically active biofilms. QCMs have a variety of uses in addition to monitoring biofilm growth:

•Chemical and biological sensors

•Electropolymerization

•Li+  intercalation

•Corrosion studies

•Electrodeposition

Gamry Instruments would like to acknowledge Dr.Jerome T. Babauta and Professor Haluk Beyenal of the Biofilms Research Group at Washington State University for the generation of these data.

Application Note Rev. 1.0 1/11/2016 Ó Copyright 2016 Gamry Instruments, Inc.



相关文章

财报|实现订单量双位数增长,赛多利斯2024财年营收33.81亿欧元

2025年2月17日,赛多利斯发布2024年财报。2024年,赛多利斯销售收入达33.81亿欧元,去年同期为33.96亿欧元,基本持平;2024年度相关净利润为2.8亿欧元,较去年同期3.39亿欧元略......

20242025梅特勒托利多实验室综合样本

梅特勒托利多2024-2025实验室综合样本,集中展示了其在实验室设备领域的技术和产品。该样本涵盖了天平、滴定仪、pH计与电导率仪、密度计与折光率仪、热值分析仪、紫外可见分光光度计、移液器、热分析仪和......

2024年奥豪斯新品推荐填表参与抽奖互动

2024年奥豪斯推出系列新品,为实验室在通用设备的选择上提供了卓越且高效的新方案。新品包括PH电极、磁力搅拌器、大称量天平、MB系列水分仪、3万转高速冷冻离心机、摇床等,以下为您逐一介绍。PH电极新品......

天平新品专栏|如何应对多通道数据传输

我需要把称量数据传到到电脑里做后续的分析;我需要把称量数据传输到我们的LIMS系统里;我需要将称量数据传输到我们的MES系统里;手工记录麻烦且容易出错,随着智能自动化的发展,越来越多的用户对天平有数据......

天平新品专栏|如何应对批量样品称量数据

不惧繁琐,轻松数据管理批量样品称量数据如何管理?手工抄录,费时费力,容易出错;纸质打印,誊抄录写,繁琐至极;梅特勒托利多新一代高级天平新增称量结果记事本功能,助您简化批量称量工作!称量结果记事本,针对......

天平新品专栏|如何应对腐蚀性样品称量

无惧严苛环境,称量随心所欲!在日常称量工作中,您是否总被下面的问题困扰:天平又锈又脏,难以清洁;2.样品具有强腐蚀性,天平用了不到一年,天平表面被腐蚀,或者溅落的液体将接口腐蚀。3.称量液体频繁,经常......

天平新品专栏|如何应对复杂环境下称量

在称量中您是否遇到这样的困扰?通风柜或层流台中的称量迟迟不能稳定,难以捕获稳定值,称量重复性差?在空间有限或者需要频繁称量的时候,1mg常规带有防风罩的天平无法摆放或操作繁杂?应采用何种措施来优化称量......

清华大学科学博物馆举行天美公司历史科学仪器捐赠仪式

2024年4月3日,天美公司向清华大学科学博物馆捐赠历史科学仪器的仪式在清华大学蒙民伟人文楼举行。天美公司历史科学仪器捐赠仪式现场本次捐赠仪式由清华大学科学博物馆收藏部负责人、科学史系助理教授刘年凯主......

称量仪器的日常测试

“应定期[⋯]依照国际和/或国内可追溯的测量标准对测量设备进行校准或验证。”——ISO9001:2008,7.6控制监视和测量设备“应当设计适用于评估设备性能的书面计划,定期校准、检测或检查自动、机械......

天平使用概念详解——最小样品重量

正确的做法是确定各称量过程的准确度(允差)要求,即,应用和法规要求的目标称量值的最大允许偏差。如图2所示,红色区中的称量会产生不正确的称量值,因为该区的仪器测量不确定度大于称量过程所需的准确度。因此,......